Troubleshooting Process Chart
- Probable Instrument Fault
- Correct / Change Set-up & Re-run Analysis
- Probable Methodology Fault
- Leak in System - Pump Fault
- Detector Fault or Injector Fault
- Pump Fault or Column Compartment Fault
- Autosampler Fault
- Improve Method
- High Back Pressure Issues
- Run a Known Method on the System
- Look for Differences Between Systems
- Probable Instrument Fault
- Correct / Change Set-up & Re-run Analysis
- Probable Methodology Fault
- Leak in System - Pump Fault
- Detector Fault or Injector Fault
- Pump Fault or Column Compartment Fault
- Autosampler Fault
- Improve Method
- High Back Pressure Issues
- Run a Known Method on the System
- Look for Differences Between Systems
Step 1 – Do Nothing!
This is your chance to assess the system as it was set up – don’t destroy the evidence.
- Are vials in the correct position?
- Is the mobile phase correct and on the correct lines?
- Error messages on the system?
- Is the correct column on the system? (Is it connected correctly?)
- Is there evidence of a leak?